Specifications
General specifications | Mounting method | Panel mount |
---|---|---|
Dimensions | 300 (W) × 300 (H) × 221 (D) mm | |
Power supply voltage | 100–240 V AC, 50/60 Hz | |
Power consumption | Approx. 80 VA (at 240 V AC) | |
Terminal | M3 screw terminal (Power terminal is M4 screw) | |
Input section | Number of inputs | 9, 18, 27, or 36 (as specified by order) |
Measurement cycle | 9 or 18 inputs: 100 ms
27 or 36 inputs: 200 ms |
|
Input signal | Thermocouple (type B, R, S, K, E, J, T, N, W, L, U, or PN)
RTD (Pt100) DC voltage (50 mV, 500 mV, or 5 V) DC current (with shunt resister) |
|
Burnout function | Provided as standard for thermocouple or RTD input | |
Display unit | Display | 12" TFT color LCD (800 × 600 dots) with automatic display off function |
Life of backlight | 50,000 hours | |
Display contents | Trend, bar graph, digital display, event summery, tag,
Historical trend, parameter, parameter setting, group, analog meter, Scale, total data, Ethernet log |
|
Recording function | Storage medium | CompactFlash |
Recording capacity | 1 GB max. | |
Data saving cycle | Links to refreshment cycle of the trend display | |
Trend data | Average, instantaneous value, or maximum and minimum value are saved. | |
Event data | Alarm data, message data, power on and off are saved. | |
Data format | ASCII (readable by Excel)
Binary (not readable by Excel) |
|
Storage capacity | Approx. 1.5 years' worth at a display refresh cycle of 30 seconds (ASCII, 9 channels, 256 MB CompactFlash) | |
Memory consumption | Displayed in bar graph
Recording stops when the memory is full. |
|
Calculation function | Difference between channels, F value, integration, square root extraction | |
Alarm | Number of settings | Up to 4 alarms for each channel |
Alarm type | High/low limits | |
Indication | Alarm status is displayed on digital display when an alarm occurs.
Histories are displayed in the alarm summary. |
|
Relay contact output | up to 20 points (option) | |
Open collector output | 16 points (option) | |
Performance | Indication accuracy | ±(0.15% + 1 digit) of input range |
Indication resolution | 0.1°C | |
Reference junction
compensation error |
±0.5°C | |
Others | Memory backup | Parameter settings are saved to the internal nonvolatile memory.
The clock is backed up by a built-in lithium battery. Trend data is stored in CompactFlash with no back-up. |
Option | Alarm output | Relay output: 10 or 20 points
Open collector output: 16 points |
Digital input | No-voltage contact input, 16 points | |
Communication | Ethernet | |
Accessories | CD-ROM (parameter loader software, instruction manual), panel mounting bracket, noise filter for power cable |
Recorders
Specifications︰
Large size paperless recorder
Front panel 300 x 300 mm
Up to 36 channels for thermocouples, RTD, voltage, or current input
Data saving on CompactFlash
For example, the data of 3 years' worth can be stored on 512 MB card.
Data viewer and parameter loader software provided as standard
Ethernet (option)
allows you to use FTP, web server, e-mail, etc.
Quick search and display of past data
Just scroll the screen to check the past data
9, 18, 27, or 36 input channels
Min Order︰ 1
Ship Date︰ 3 DAYS